Abbott Glucose Test
Strips: Recall - False Low Blood Glucose
Results
including Precision Xceed Pro, Precision Xtra, Medisense
Optium, Optium, OptiumEZ, ReliOn Ultima
AUDIENCE: Endocrine, Patients,
Pharmacy
ISSUE: FDA and Abbott Diabetes Care
notified healthcare professionals and patients of a recall of
359 different lots of glucose test strips marketed under the
following brand names: Precision Xceed Pro, Precision Xtra,
Medisense Optium, Optium, OptiumEZ and ReliOn Ultima. The
problem relates to a defect that inhibits sufficient absorption
of blood into the test strip. Strips exposed to warm weather or
prolonged storage may be more likely to provide a false result.
Test strips with lot numbers that have been recalled may give
falsely low blood glucose results, which can lead patients to
try to raise their blood glucose when it is unnecessary, or to
fail to treat elevated blood glucose due to a falsely low
reading. Both scenarios pose risks to health.
BACKGROUND: These strips are used with
Abbott’s Precision Xtra, Precision Xceed Pro, MediSense Optium,
Optium, Optium EZ and ReliOn Ultima blood glucose monitoring
systems. As many as 359 million strips may be affected by the
recall. The test strips, which were manufactured between
January and May 2010, are sold both in retail and online
settings directly to consumers, but are also used in health
care facilities.
RECOMMENDATION: FDA is releasing
recommendations for consumers and healthcare professionals.
These recommendations explain how to determine whether a
particular lot is affected, how to order a free replacement set
of strips, and what steps to take in the meantime.
Patients with diabetes should be aware of this
problem and take steps to prevent it from affecting their
health. Customers can check if they have tests trips from
the recalled lots by visiting Abbott’s website to look up
their product lot number: http://www.precisionoptiuminfo.com
Healthcare professionals and patients are
encouraged to report adverse events or side effects related to
the use of these products to the FDA's MedWatch Safety
Information and Adverse Event Reporting
Program:
- Complete and submit the
report Online: www.fda.gov/MedWatch/report.htm
- Download
form or call
1-800-332-1088 to request a reporting form, then
complete and return to the address on the
pre-addressed form, or submit by fax to
1-800-FDA-0178
Read the MedWatch safety alert,
including links to the News Release and FDA Questions and
Answers, at:
http://www.fda.gov/Safety/MedWatch/SafetyInformation/SafetyAlertsforHumanMedicalProducts/ucm237910.htm
Tell us how we are
doing:
http://survey.foreseeresults.com/survey/display?cid=d509U9xg0gZVxkMRcoMsQg==&sid=link&cpp[date]=12_22_2010_1240&cpp[type]=S
You are encouraged to report all serious
adverse events and product quality problems to FDA MedWatch
at
www.fda.gov/medwatch/report.htm
NOTE: The Division of EMS
does not endorse products or services. The contents of
this e-mail are provided as a service to the EMS
community.
|